Clean Skin & Scalp Surface
First) Use Lemon Prep Abrasive Scrub and/or Alcohol Wipe to remove oils, makeup, etc from the surface; this can be done using gauze, Q-tip, glove, or finger. Be Careful to ask about Lemon Allergies!
Next) Remove the Lemon Prep with a tissue, gauze, or paper towel. Add water if needed. Be sure to use non-lotion infused tissues!
Apply Electrodes to Clean Surface
Place the electrodes in their appropriate locations.
ERG = 2 Corneal electrodes, 2 Temple electrodes, 1 Forehead electrode
VEP = 1 Forehead electrode, 1 Crown electrode, 1 Occipital electrode
EOG = 1 electrode on each side of both eyes (4 total), 1 Forehead electrode
Plug Electrodes Into Amplifier
ERG
Left Eye: (-) Temple (+) Corneal
Right Eye: (-) Temple (+) Corneal
Forehead: Green GND / Cz
VEP
Fz: Forehead
Cz: Crown
Oz: Occipital
O1 & O2: Occipital for 3 channel pVEP
EOG
Left Eye: (-) outer (+) inner eye
Right Eye: (-) outer (+) inner eye
Forehead: Green GND / Cz
Pupillometry, FST, and DA do not require electrode placement.