Message: Color Matrix for Stimulator 1, Serial Number ### cannot be found. Using Default Matrix
Cause: Calibration files not present
Action 1: Check for calibration files; are calibration files present in the software folder. To check for these files, open Window's start
menu -> Computer -> C: drive
. Choose the Multifocal
or Espion
folder. Search this folder for a .CAL
file. It should look like COLD###.CAL
.
Action 2: Check the software configuration. Is the stimulator serial number listed? Check for the serial number in Configure System -> Hardware
-> Ganzfeld -> Serial Number
. The serial number should match both the stimulator's decal as well as the number of the calibration file.
Action 3: Check that IO card
is enabled
. IO Card Enabled
can be found in Configure System -> Hardware -> Console
.
Definition:
.CAL is a file extension used for calibration files.
Calibration files ensure the stimulator is firing accurate increments of light.
Configure System is a place to specify all system parameters.